Any questions? Call 310-318-1500
Dr.UGraft™ punch which arrives in a travel-ready suitcase comes with two components that require sterilization upon first use. This includes the hand piece attached to the power cord, and secondly, the punch tip itself.
The pre-packaged tubing has already been sterilized. But as noted earlier, check to be sure that the package seal has not been broken. Otherwise it will need to be discarded, as opposed to using or sterilizing the tubing.
The Dr.UGraft hand piece and punch will need to be sterilized for the first time after they arrive.The Dr.UGraft ™ Nurture handpiece arrives with the power cord already attached as a single unit. And both must be sterilized together. They are never to be detached for any reason.
Here is the process to sterilize the Dr.UGraft™ Nurture hand piece.
Following the guidelines provided above (i.e. 134-135 degrees Celcius for 15 minutes) is one way to ensure sterilization. Likewise, you can also refer to the color-change indicator on the bag itself. If the arrow stays pink, this means that the sterilization process has not reached completion. The color brown indicates that the contents have been properly sterilized.
Another method for sterilizing the Dr.UGraft ™ hand piece and punch is the method of using ionized gas, also known as plasma sterilization.
The actual hybrid, flared punch unit must be sterilized in a separate, appropriate-sized bag. Once sealed, place the bag inside the autoclave machine, following the standards provided above (134-135 C for 15 min). You can check the color indicator on the outside of the pouch to make sure that the arrow has changed from pink to brown, as confirmation that the punch has been thoroughly sterilized.
Next, learn how to assemble all the Dr.UGraft ™ components together into a comprehensive working system
Read the complete Dr.UGraft ™ online instructional guide here.